| Model Number |
701118090 |
| Health Effect - Clinical Code |
Needle Stick/Puncture (2462)
|
| Date of Event |
10/08/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The sample is indicated as available for return.As of the date of this report, the sample has not been returned.A follow-up report will be provided once the device has been received and reviewed.
|
| |
|
Event or Problem Description
|
|
The supercore needle fell apart during use.The doctor stabbed his hand with the broken device, left hand, little finger.No injury to patient.
|
| |
|
Additional Manufacturer Narrative
|
|
One opened device was returned for evaluation.Visual inspection found that the stylet was separated from the body of the device, with signs of the body having split open proximally.The complaint was confirmed.Argon medical devices has received other complaints that the supercore semi-automatic biopsy instrument is coming apart during or prior to use.Argon has conducted an internal investigation and tracked the affected parts to a narrow time frame resulting from a specific manufacturing event.The plastic housing and plunger can be separated more easily than normal for the lots manufactured during this time frame.Capa 2021-061 has been initiated to document our investigation into the cause of this problem and the corrective action that are being taken to prevent the re-occurrence of this problem.To ensure continued customer satisfaction, argon medical devices has decided to issue a voluntary recall of the affected lots because of the high rate of reports of unintentional disassembly of these devices.
|
| |
|
Event or Problem Description
|
|
The supercore needle fell apart during use.The doctor stabbed his hand with the broken device, left hand, little finger.No injury to patient.
|
| |
|