| Model Number |
172516-LVISJ-C-PMA |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/20/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformances associated with the reported part/lot number combination did not reveal any production-relatedissues relevant to the complaint that occurred during manufacturing of the device.The reported complaint is unconfirmed as the stent was able to fully open upon deployment during the investigation.The distortion of the proximal long loops is consistent with the stent being retrieved from the vessel with a snare during the procedure.The distal long loops were damaged while loading the stent during the investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that after deploying the lvis jr.Stent, the back half of the stent leading to the proximal end did not open and appeared flattened out.The stent was removed with a snare.There was no reported patient injury.
|
| |
|
Search Alerts/Recalls
|