BOSTON SCIENTIFIC CORPORATION INOGEN MINI ICD DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number D012
Medical Device Problem Codes Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574); Unexpected Therapeutic Results (1631)
Health Effect - Clinical Code Electric Shock (2554)
Date of Event 09/05/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient with this implantable cardioverter defibrillator (icd), presented to the emergency room after getting shocks for sinus tachycardia.Review of a consult transmission exhibited inappropriate anti-tachycardia pacing (atp) due to atrial driven tachycardia episodes, with therapy exhaustion observed in an episode.Technical services (ts) also noted device battery status had reached elective replacement indicator, and recommended replacing device within 90 days.Subsequently, the device was explanted and replaced with a new device.No additional adverse patient effects were reported.At this time, the device has not been returned for analysis.If the device is received, product analysis will be completed, and this report will be updated with pertinent information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN MINI ICD DR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key12662468
Report Number2124215-2021-31557
Device Sequence Number4852056
Product Code LWS
UDI-Device Identifier00802526533822
UDI-Public00802526533822
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P960040/S306
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/25/2016
Device Model NumberD012
Device Catalogue NumberD012
Device Lot Number104426
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/27/2021
Initial Report FDA Received Date10/19/2021
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age86 YR
-
-