• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. DORADO PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. DORADO PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number DR8044
Medical Device Problem Codes Break (1069); Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/22/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one dorado pta dilatation catheter has been received for the evaluation.On the visual evaluation, the device appeared bloody and no other anomalies were noted.On the functional evaluation, an in-house presto inflation device was used to inflate the balloon but was unable to be inflated or maintain pressure.The catheter shaft was then cut multiple times and attached to a touhy adapter to be inflated but was still unable to inflate.The catheter was examined under magnification, the dried blood and contrast was noted in the catheter.The circumferential break was also noted at the proximal balloon joint.No further testing was performed.Therefore the investigation for the reported leakage remains inconclusive, as the functional testing was not performed due to the contrast medium and the blood that was noted to be dried inside the inflation lumen of the returned device.The investigation for the identified circumferential break was confirmed, as the break was noted at the proximal balloon joint under the microscopic observation.A definitive root cause for the reported leakage and the identified circumferential break could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.(expiry date: 12/2023).
 
Event or Problem Description
It was reported that during angioplasty procedure through superficial femoral artery via cfa/popliteal, the balloon allegedly had leakage.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one dorado pta dilatation catheter has been received for the evaluation.On the visual evaluation, the device appeared bloody and no other anomalies were noted.On the functional evaluation, an in-house presto inflation device was used to inflate the balloon but was unable to be inflated or maintain pressure.The catheter shaft was then cut multiple times and attached to a touhy adapter to be inflated but was still unable to inflate.The catheter was examined under magnification, the dried blood and contrast was noted in the catheter.The circumferential break was also noted at the proximal balloon joint.No further testing was performed.Therefore the investigation for the reported leakage remains inconclusive, as the functional testing was not performed due to the contrast medium and the blood that was noted to be dried inside the inflation lumen of the returned device.The investigation for the identified circumferential break was confirmed, as the break was noted at the proximal balloon joint under the microscopic observation.The identified circumferential break is likely the root cause for the reported leakage.However, the definitive root cause for the reported leakage and the identified circumferential break could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 12/2023), g3.H11: e1(initial reporter city).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Event or Problem Description
It was reported that during an angioplasty procedure through superficial femoral artery via cfa/popliteal, the balloon allegedly had leakage.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORADO PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
FUTUREMATRIX INTERVENTIONAL
1605 enterprise street
athens 75751
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key12662812
Report Number2020394-2021-01891
Device Sequence Number4298302
Product Code LIT
UDI-Device Identifier00801741058301
UDI-Public(01)00801741058301
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K072283
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDR8044
Device Catalogue NumberDR8044
Device Lot Number93ZE0067
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/11/2021
Supplement Date Received by Manufacturer12/21/2021
Initial Report FDA Received Date10/20/2021
Supplement Report FDA Received Date12/23/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-