Model Number 8888222316
Medical Device Problem Code Difficult to Advance (2920)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/29/2021
Type of Reportable Event Death
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during procedure, when the doctor was inserting the catheter in intensive care unit (icu) to start continuous renal replacement therapy (crrt), the guide wire kept bending and it took three attempts and three different kits to place the line.Each attempt was completed with new kit, new catheter and new wire.The first kit that had the issue was replaced by a second kit with the same lot which also had the same issue.The third kit with the same item number was used on the same day of the event, immediately following unsuccessful attempts on the first two.The third kit was used successfully and the procedure was completed.It was stated that the bent/issue with the guide wire was not noticed in the packaging.There were no pieces to be retrieved.There was a resistance when inserting or withdrawing the guidewire.The guide wire used was the one included in the kit.Nothing unusual was observed on the device prior to use.No other products were being utilized with the device.No visible defect or damage was noted with the catheter or puncture needle.The catheter was not repaired, there was no leak, tego was not utilized, and there was no luer adapter issue.There was no blood loss.No additional medical or interventional treatments were provided as a result of the bent guidewire/line placement.The patient's status at the time of the event was alive but in icu.The patient did not experience any harms, symptoms or complications as a result of the reported event with the bent guidewire/line placement.The patient died as a result of illness after eventual successful placement of the acute catheter.The patient death was not related to the catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAHURKAR ELITE
Common Device Name
CATHETER, HEMODIALYSIS, NON-IMPLANTED
Manufacturer (Section D)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
CS  20101
MDR Report Key12665159
Report Number3009211636-2021-00292
Device Sequence Number16451728
Product Code MPB
UDI-Device Identifier10884521126473
UDI-Public10884521126473
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K120674
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/01/2022
Device Model Number8888222316
Device Catalogue Number8888222316
Device Lot Number2104600199
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/29/2021
Initial Report FDA Received Date10/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
-
-