|
It is reported by the customer, during a therapeutic colonoscopy using a single use loop cutter, the device malfunctioned inside the patient.The broken parts were retrieved from the patient.The device malfunction caused a clinically significant delay in the procedure.No additional consequences to the patient have been reported.Additional details regarding the patient and procedure have been reported.At this time, no additional information has been provided.
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The dhr with the lot number of the subject device was confirmed.No abnormalities detected in the dhr for the following items which related to the reported phenomenon.The exact phenomenon could not be identified.·process inspection slip.·quality inspection slip.·nonconformity processing slip.The instructions for use (ifu) instruction manual state: ¿the instruction manual contains the following descriptions, and it warns against this event.¿before use, prepare and inspect the instrument as instructed below.Should any irregularity be observed, do not use the instrument; use a spare instead.Damage or irregularity may compromise patient or user safety, such as punctures, hemorrhages, or mucous membrane damage, and may result in more severe equipment damage.¿never use excessive force to open or close the cutter.This could damage the instrument.¿do not insert the instrument into the endoscope unless you have a clear endoscopic field of view.If you cannot see the distal end of the insertion portion in the endoscopic field of view, do not use it.This could cause patient injury, such as punctures, hemorrhages, or mucous membrane damage.It may also damage the endoscope and/or instrument.¿do not angulate the bending section of the endoscope while the distal end of the insertion portion is extended from the distal end of the endoscope.This could cause patient injury, such as punctures, hemorrhages, or mucous membrane damage.¿when inserting the instrument into the endoscope, hold the slider firmly.Otherwise, the cutter may open and extend from the endoscope tip abruptly.This could cause patient injury, such as punctures, hemorrhages, or mucous membrane damage.It could also damage the endoscope and/or instrument.¿do not withdraw the instrument from the endoscope while the cutter is open.This could damage the endoscope and/or instrument.¿if excessive resistance makes withdrawal difficult, adjust the angle of the endoscope until the instrument can be withdrawn smoothly.Forcible withdrawal could damage the instrument and/or endoscope.¿ olympus will continue to monitor the field performance of this device.
|