• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC VIPER2 T20 HEXLOBE DRIVER SHAF; DRIVER, PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC VIPER2 T20 HEXLOBE DRIVER SHAF; DRIVER, PROSTHESIS Back to Search Results
Model Number 286725020
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If the information is unknown, not available or does not apply, the section/field of the form is left blank.Event occurred on an unknown date in 2021.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date during an anterior/posterior lumbar fusion procedure, , one (1) viper universal poly driver was broken, and two (2) t20 driver shafts were stripped.The surgeon was placing unknown screws in the patient when two (2) poly screwdriver shafts tips broke off.He then attempted to use the viper2 t20 hexlobe driver and t20 screwdriver shaft which both stripped.He advanced using the viper universal poly driver and the tip of the driver was found to be broken on the back table after it was taken out.There were no fragments generated.There was no patient consequence or surgical delay.The procedure was successfully completed.This report is for one (1) viper2 t20 hexlobe driver shaft.This is report 1 of 5 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER2 T20 HEXLOBE DRIVER SHAF
Common Device Name
DRIVER, PROSTHESIS
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
MDR Report Key12665928
Report Number1526439-2021-02220
Device Sequence Number16439358
Product Code HWR
UDI-Device Identifier10705034214290
UDI-Public10705034214290
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number286725020
Device Catalogue Number286725020
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/23/2021
Initial Report FDA Received Date10/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
POLY SCW DRIVER SHAFT, CANNULTD; POLY SCW DRIVER SHAFT, CANNULTD; T20 SCREWDRIVER SHAFT; UNKNOWN AWLS; UNKNOWN SCREWS; VIPER UNIVERSAL POLY DRIVER
-
-