Model Number A303314A
Medical Device Problem Code Illegible Information (4050)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the customer had an issue with some bard foley kits where the foleys were not properly labeled with the latex warning sticker.It was stated that vet had order for a foley placement, allergies verified no allergy to latex noted and original ordered read to place a 16f foley and if not successful to place a 14f foley.It was stated that when inspecting the foley kits cather kits reference numbers a303316a for 16 french foley cath kit and a303314a 14 french foley cath kit, one had a label with a warning for latex and the other did not.It was also stated that the warning label should be documented on all kits and the actual product warning was hard to see as it was on the seem of the product (material: a303316a and a303314a).It was also stated that if this was a patient who had an allergy they could be harmed as there were no kits without latex.
 
Event or Problem Description
It was reported that the customer had an issue with some bard foley kits where the foleys were not properly labeled with the latex warning sticker.It was stated that vet had order for a foley placement, allergies verified no allergy to latex noted and original ordered read to place a 16f foley and if not successful to place a 14f foley.It was stated that when inspecting the foley kits cather kits reference numbers a303316a for 16 french foley cath kit and a303314a 14 french foley cath kit, one had a label with a warning for latex and the other did not.It was also stated that the warning label should be documented on all kits and the actual product warning was hard to see as it was on the seem of the product (material: a303316a and a303314a).It was also stated that if this was a patient who had an allergy they could be harmed as there were no kits without latex.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be ¿vendor/ printer error".It is unknown whether the device had met relevant specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "sterile unless package is opened or damaged, except for any individually packaged components within the tray which are not labeled as sterile.These components are not terminally sterilized.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Single use only.Do not resterilize.For urological use only.Warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage the catheter and may cause balloon to burst.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practices and applicable local, state and federal laws and regulations.Visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.Please consult product label and insert for any indications, contraindications, hazards, warnings, cautions and directions for use.Proper techniques for urinary catheter insertion.Perform hand hygiene immediately before and after insertion.Insert urinary catheters using aseptic technique and sterile equipment.Use the smallest foley catheter possible, consistent with good drainage.Document the indications for catheter insertion, date and time of catheter insertion, individual who inserted catheter, and date and time of catheter removal in patient.Record.Proper techniques for urinary catheter maintenance.Secure the foley catheter, use the statlock® foley stabilization device if provided.Maintain a closed drainage system by utilizing pre-connected, sealed.Catheter-tubing junctions.Maintain unobstructed urine flow and keep the catheter and collection tube.Free from kinking.Keep the collection bag below the level of the bladder or hips at all times.Empty the collection bag regularly (e.G., prior to transport) using a separate, clean collection container for each patient.Routine hygiene (e.G., cleansing of the meatal surface during daily bathing.Or showering) is appropriate.Leave foley catheter in place only as long as needed." h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESTEP¿ FOLEY TRAY SYSTEM BARDEX® IC COMPLETE CARE®
Common Device Name
FOLEY TRAY
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key12666264
Report Number1018233-2021-06505
Device Sequence Number20046718
Product Code MJC
UDI-Device Identifier00801741073786
UDI-Public(01)00801741073786
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K910318
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA303314A
Device Catalogue NumberA303314A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/29/2021
Supplement Date Received by Manufacturer03/29/2022
Initial Report FDA Received Date10/20/2021
Supplement Report FDA Received Date04/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-