| Model Number |
IPN913776 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
|
| |
|
Event or Problem Description
|
|
It was reported the unit is "not powering on".No patient involvement reported.
|
| |
|
Event or Problem Description
|
|
It was reported the unit is "not powering on".No patient involvement reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).The sample was returned for evaluation.A visual exam was performed and there were no major defects observed.Functional testing was performed and the unit was connected to 120vac; however, the unit did not pass the power-on self-test (p.O.S.T.).The power button was not illuminating and when pressed , the unit did not power on.The unit was opened and it was noted that the power harness had heat damage at its connector.The power supply board was replaced with a known good lab inventory board.This time, the unit was able to pass the power-on self-test with no issues.During the temperature display accuracy test the unit displayed the correct temperatures and properly alarmed in the high temperature scenario.The complaint has been confirmed.The investigation revealed a defective power supply board.Since the neptunes are 100% functionally tested during manufacturing assembly, it is unlikely that the failure was present at the time of release.A device history record review was performed with no evidence to suggest a manufacturing related issue.The unit was manufactured in 2015 and was designed for a minimum of 5 years.The root cause for the failure is normal wear.Teleflex will continue to monitor and trend on complaints of this nature.
|
| |
|
Search Alerts/Recalls
|