BOSTON SCIENTIFIC CORPORATION V-14 CONTROL WIRE; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 40458
Medical Device Problem Codes Detachment of Device or Device Component (2907); Difficult to Advance (2920); Material Twisted/Bent (2981)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 09/23/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that guidewire tip detachment occurred.The patient presented with atherosclerotic disease in tibial arteries.The target lesion was located in the tibial artery.During a retrograde procedure, a 300cm v-14 control guidewire was advanced and a 018 non-boston scientific catheter was used to support the wire.However, there was lack of wire traction and the wire looped.Subsequently, 5mm of the guidewire tip detached.Snaring of the detached fragment was attempted and failed.The physician then used a balloon expandable stent to jail the tip of the guidewire in the tibial vessel.The procedure was completed and no further patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-14 CONTROL WIRE
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key12669346
Report Number2134265-2021-12737
Device Sequence Number19874122
Product Code DQX
UDI-Device Identifier08714729799269
UDI-Public08714729799269
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K112745
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/18/2022
Device Model Number40458
Device Catalogue Number40458
Device Lot Number0026042676
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/30/2021
Initial Report FDA Received Date10/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-