Model Number N/A
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/22/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Concomitant devices -tibia cemented 5 degree stemmed right size g catalog # 42532007902 lot #: 64831313, articular surface fixed bearing ultracongruent (uc) right 14 mm height catalog # 42522200614 lot #: 63722485.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this patient; please see all reports associated with this event: 3007963827 - 2021 - 00245.Investigation incomplete.
 
Event or Problem Description
It was reported that after the tibia and femur were cemented, the surgeon felt that there was an issue with the anterior portion of the tibia where the poly insertion instrument engages for final poly insertion.The instrument would not engage correctly and the poly wouldn't seat.Attempts have been made and no further information has been provided.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up is being submitted to relay additional information.The results of the investigation are as follows: -no product was returned or pictures provided; visual and dimensional evaluations could not be performed.- review of the device history record(s) identified no deviations or anomalies during manufacturing.H results as applicable.-a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA ARTICULAR SURFACE FIXED BEARING(UC)RIGHT 14MM HEIGHT
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12669565
Report Number3007963827-2021-00244
Device Sequence Number7512474
Product Code MBH
UDI-Device Identifier00889024243156
UDI-Public(01)00889024243156(17)220731(10)63722485
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K121771
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2022
Device Model NumberN/A
Device Catalogue Number42522200614
Device Lot Number63722485
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2021
Supplement Date Received by Manufacturer02/09/2022
Initial Report FDA Received Date10/20/2021
Supplement Report FDA Received Date02/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/20/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE.
Patient SexMale
Patient Weight78 KG
-
-