• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET UK LTD. 32MM MOD HEAD COCR -6MM NECK; ZIMMER BIOMET TYPE1 COCR HEADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. 32MM MOD HEAD COCR -6MM NECK; ZIMMER BIOMET TYPE1 COCR HEADS Back to Search Results
Model Number N/A
Medical Device Problem Codes Material Integrity Problem (2978); Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Pain (1994); Metal Related Pathology (4530)
Date of Event 09/30/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Initial report.Report source, foreign - event occurred in (b)(6).Concomitant medical devices: customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Associated products: medical product: taperloc bm/pc lat 10.0x140 t1, catalog no.: 11-103204bm, lot no.: 6423447.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that a patient underwent an initial hi arthroplasty on (b)(6) 2019.Subsequently, a revision procedure due to metallosis was performed on (b)(6) 2021.Surgeon, mr (b)(6) was concerned about the severe amount of metallosis and thought it may be related wear of the femoral head and stem implants.
 
Additional Manufacturer Narrative
(b)(4).This final report is being submitted to relay additional information.Complaint summary: no product was returned, and no pictures provided; visual and dimensional evaluations could not be performed.A review of the device history records (dhr j6250944) have been checked and verify that the parts were manufactured and sterilised in accordance with the applicable specifications with no deviations or non-conformances found during manufacturing and inspection.The device is used for treatment.The reported product was reviewed for compatibility with no issues noted.A review of complaint history found no additional related issues for this item and the reported part and lot combination.It has been confirmed that the item is not within the scope or subject of any field actions or recalls which could be attributed to reported event.Medical records/radiographs were not provided.A definitive root cause cannot be determined.The likely condition of the devices when they left zimmer biomet is conforming to specification.The reported event has not been confirmed as the products have not been returned for evaluation and the dhr review did not identify any issues.The root cause of the reported event cannot be determined with the information provided.No corrective or preventive actions resulted after the investigation of this event.The investigation is completed based on current available information.If any additional information becomes available, then the complaint will be reopened and further investigated.H3 other text: item not returned.
 
Event or Problem Description
It was reported that a patient underwent an initial hi arthroplasty on (b)(6) 2019.Subsequently, a revision procedure due to metallosis was performed on (b)(6) 2021.Surgeon, mr.(b)(6) was concerned about the severe amount of metallosis and thought it may be related wear of the femoral head and stem implants.
 
Additional Manufacturer Narrative
(b)(4).This follow-up final report is being submitted to relay additional information as the product has been returned for evaluation.A dimensional inspection was conducted and the the 32mm modular head cocr -6mm neck item:163667 lot:j6250944 was found to be conforming to specification.The likely condition of the device when it left zimmer biomet is conforming to specification.The reported event has not been confirmed as the dimensional inspection shows the product is conforming to specification and the dhr review did not identify any issues.The root cause of the reported event cannot be determined with the information provided.
 
Event or Problem Description
It was reported that a patient underwent an initial hip arthroplasty.Subsequently, a revision procedure due to metallosis was performed approximately 2 years, 4 months later.Surgeon, mr (b)(6) was concerned about the severe amount of metallosis and thought it may be related wear of the femoral head and stem implants.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
(b)(4).This follow-up final report is being submitted to relay additional information.A ø32 mm cocrmo femoral head was received for analysis, due to revision after approximately 2 years and 4 months in service.Visual examination of the cocrmo head shows fine scratches on the bearing surface, which may have been caused by third body debris, however this cannot be confirmed without provision of radiographs and revision surgical notes.The taper surface of the cocrmo head shows some discolouration which may indicate some fretting has occurred.The root cause of the of the reported alval and metallosis could not be determined in this instance.Potential contributing factors such as the presence of third body debris within the joint space and/or component positioning cannot be discussed without provision of post-primary and pre-revision full pelvis radiographs, primary and revision surgery notes, and additional patient information.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
32MM MOD HEAD COCR -6MM NECK
Common Device Name
ZIMMER BIOMET TYPE1 COCR HEADS
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12670885
Report Number3002806535-2021-00452
Device Sequence Number4211885
Product Code JDL
UDI-Device Identifier00887868308314
UDI-Public00887868308314
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K042841
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 08/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number163667
Device Lot NumberJ6250944
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/07/2021
Supplement Date Received by Manufacturer01/17/2022
03/30/2022
07/29/2022
Initial Report FDA Received Date10/21/2021
Supplement Report FDA Received Date01/18/2022
03/31/2022
08/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;