• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE I PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE I PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-45
Device Problems Retraction Problem (1536); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/27/2021
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device: a supplemental report will be submitted when additional information is provided.
 
Event Description
It was reported that during a routine check the cardiosave intra-aortic balloon pump (iabp) had a stuck power cord that cannot be retracted or stretched normally.There was no patient involvement, and no adverse event was reported.
 
Manufacturer Narrative
The iabp unit was cleared for use and returned to the customer.A supplemental report will be submitted upon completion of our investigation.
 
Event Description
N/a.
 
Manufacturer Narrative
Updated fields: b4, g4, g7, h2, h10, h11.Corrected fields: a1, h6 (component codes), h10.A getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the iabp unit and rewound the power cord winch.Once completed, the fse noted that the issue was solved.A supplemental report will be submitted upon completion of our investigation.
 
Event Description
N/a.
 
Event Description
N/a.
 
Manufacturer Narrative
Updated fields: b4, g4, g7, g8, h2, h4, h6(investigation type, investigation findings & investigation conclusions), h10, h11.Corrected fields: g1(contact person).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE I PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12672544
MDR Text Key277721387
Report Number2249723-2021-02401
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108421
UDI-Public10607567108421
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup,Followup
Report Date 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-45
Device Catalogue Number0998-00-0800-45
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/27/2021
Initial Date FDA Received10/21/2021
Supplement Dates Manufacturer Received10/21/2021
11/18/2021
08/31/2022
Supplement Dates FDA Received11/17/2021
12/14/2021
09/02/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/28/2020
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-