DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE I PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-45 |
Device Problems
Retraction Problem (1536); Mechanical Jam (2983)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/27/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Testing of actual/suspected device: a supplemental report will be submitted when additional information is provided.
|
|
Event Description
|
It was reported that during a routine check the cardiosave intra-aortic balloon pump (iabp) had a stuck power cord that cannot be retracted or stretched normally.There was no patient involvement, and no adverse event was reported.
|
|
Manufacturer Narrative
|
The iabp unit was cleared for use and returned to the customer.A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
Updated fields: b4, g4, g7, h2, h10, h11.Corrected fields: a1, h6 (component codes), h10.A getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the iabp unit and rewound the power cord winch.Once completed, the fse noted that the issue was solved.A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
N/a.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
Updated fields: b4, g4, g7, g8, h2, h4, h6(investigation type, investigation findings & investigation conclusions), h10, h11.Corrected fields: g1(contact person).
|
|