• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. INDIGO SYSTEM ASPIRATION CATHETER 6; QEW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. INDIGO SYSTEM ASPIRATION CATHETER 6; QEW Back to Search Results
Model Number CAT6KIT-A
Medical Device Problem Codes Material Deformation (2976); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/27/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2021-02345.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure in the left superficial femoral artery (sfa) and popliteal artery using indigo system aspiration catheter 6s (cat6), a non-penumbra catheter, a non-penumbra sheath, and a guidewire.During the procedure, the physician aspirated some clot using the non-penumbra catheter.Then, the physician advanced a cat6 to the target vessel through the sheath.It is unknown if a peel away sheath was used with the cat6.Subsequently, the physician aspirated and removed clot using the cat6.Upon removal of the cat6, the physician noticed the cat6 was kinked at the distal end.It was also reported that the physician flushed and wiped the cat6 with saline prior to insertion and that the cat6 was not advancing as well as normal.Next, while advancing another cat6 to the target vessel through the sheath using the peel away sheath, the physician experienced resistance.Subsequently, the physician kinked the cat6 at the distal end.Therefore, the cat6 was removed.It was reported that the physician flushed and wiped the cat6 with saline prior to insertion.The procedure was completed using the previously used non-penumbra catheter and another non-penumbra sheath.It was reported that the patient was taken to surgery for a femoral-popliteal bypass graft placement due to the non-penumbra catheter perforating the popliteal artery.There was no report of an adverse effect to the patient related to the use of the indigo aspiration system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INDIGO SYSTEM ASPIRATION CATHETER 6
Common Device Name
QEW
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key12675173
Report Number3005168196-2021-02346
Device Sequence Number1181347
Product Code QEW
UDI-Device Identifier00815948020405
UDI-Public00815948020405
Combination Product (Y/N)Y
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K192833
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/21/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/14/2024
Device Model NumberCAT6KIT-A
Device Catalogue NumberCAT6KIT
Device Lot NumberF107084
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/27/2021
Initial Report FDA Received Date10/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age71 YR
Date Report Sent to FDA01/01/2005
-
-