Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.A review of the manufacturing records was performed, and no non-conformances were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Investigation conclusion: a complaint lot history check was performed on lot # 0013357 for needles bent.This is the 1st related complaint for needles bent on lot # 0013357.A review of the device history record was completed for batch# 0013357.All inspections were performed per the applicable operations qc specification.There were zero (0) notifications noted that pertained to the complaint.As no samples and/or photo(s) were received the investigation concluded: unconfirmed: bd was not able to duplicate or confirm the customer¿s indicated failure as no samples or photos were returned root cause cannot be determined at this time as the issue is unconfirmed as no samples or photos were returned.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.
|
It was reported that the bd 1ml 29g syringe experienced product damage while still considered operable.The following information was provided by the initial reporter, translated from chinese to english: on september 3, diabetic patients bought disposable sterile insulin syringes, and found that the syringes were bent on the head when they opened the package, so they stopped using this product.No patient information.
|