MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRNY II CR LOCK FEM IMPLANT IMPACTOR; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRNY II CR LOCK FEM IMPLANT IMPACTOR; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 74011711
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/04/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that the plastic locking ring of a journey ii cr locking femoral implant impactor is broken.As this was noticed upon field inspection, there was no patient involvement.
 
Additional Manufacturer Narrative
The device, intended for use in treatment was returned for evaluation: a visual inspection of the device confirms the knob ring is broken which causes it to not properly function.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incidents.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II CR LOCK FEM IMPLANT IMPACTOR
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12677305
Report Number1020279-2021-07644
Device Sequence Number1454448
Product Code JWH
UDI-Device Identifier00885556422892
UDI-Public00885556422892
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K121443
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number74011711
Device Catalogue Number74011711
Device Lot Number16BBG0021
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2021
Initial Date Received by Manufacturer 10/04/2021
Supplement Date Received by Manufacturer11/02/2021
Initial Report FDA Received Date10/21/2021
Supplement Report FDA Received Date11/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-