| Catalog Number |
07.02077.001 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/14/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Inspection the product was not returned and no photos were provided.Device evaluation could not be completed.Dhr review the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left zimmer biomet¿s control.Potential root cause possible contributors to this event could be dense bone quality of the patient, off-axis use of the probe, or attempting to alter the trajectory of the probe within the vertebrae.Since the complaint was reported outside of a case and no event was reported about how the device bent, the cause cannot be conclusively determined.Device usage this device is used for treatment.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of a straight lumbar lenke probe snapped off while trying to remove it from the patient's bone; the patient had very hard bone and the probe was difficult to remove.The tip was recovered and there was no reported patient harm.
|
| |
|