| Model Number |
TRM-485 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/14/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "left side rupture".Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: visual analysis of the returned device identified: weight to the spec, opening, red particles on the surface of the shell.A microscopic analysis was performed which identify sharp edge and with non-penetrating nicks assessed as surgical impact opening.A dimension measurement in the shell was performed which identify the thickness within specification.Non penetrating nicks.Based on the device analysis the final assessment is a sharp edge opening with non-penetrating nicks assessed as surgical impact and non-penetrating nicks.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|