CORDIS CORPORATION TRAPEASE PVCF FEM/JUG 55CM CSI; FILTER, INTRAVASCULAR, CARDIOVASCULAR
|
Back to Search Results |
|
Catalog Number 466P306AU |
Device Problems
Fracture (1260); Unintended Movement (3026); Migration (4003)
|
Patient Problems
Perforation (2001); Perforation of Vessels (2135)
|
Event Date 09/28/2021 |
Event Type
Injury
|
Event Description
|
As reported by the legal brief, the patient underwent placement of a trapease vena cava filter.The report states that the filter subsequently malfunctioned and caused injury and damage to the patient including, but not limited to filter fracture, perforation, migration and tilt.As a direct and proximate result of these malfunctions, the patient suffered life-threatening injuries and damages, and required extensive medical care and treatment.As a further proximate result, the patient has suffered and will continue to suffer significant medical expenses, pain and suffering, and other damages.
|
|
Manufacturer Narrative
|
Occupation: other, senior counsel, litigation.Please note that the exact event date is unknown and the event date is the complaint awareness date.As reported, the patient underwent placement of a trapease vena cava filter.The indication for the filter placement was not reported.At some point after the filter implantation, the patient became aware that the filter had tilted, fractured and migrated and was associated with perforation.The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed.The trapease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the inferior vena cava (ivc) for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pe where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pe where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Without images or procedural films for review, the reported filter tilt, fracture, migration and perforation events could not be confirmed and the exact cause could not be determined.Ivc filter tilt has been associated with the anatomy of the vessel, specifically asymmetry and tortuousness.Additionally, the timing and mechanism of the filter tilt is unknown.The timing and mechanism of the fracture has not been reported at this time.The instructions for use (ifu) states that filter fracture is a potential complication of vena cava filters.Anatomic locations that create concentrated stress points from filter deformation (for example, deployment at apex of scoliosis, overlapping of either of the renal ostia, or placement adjacent to a vertebral osteophyte) may contribute to fracture of a particular filter strut.Ivc filter migration is a known potential adverse event associated with all ivc filter implants and is listed in the ifu as such.Possible causes for filter migration include mega cava, wire entrapment during central venous catheter placement, ¿sail¿ effect (cranial migration) of large clot burden within the filter, mechanical device failure, and operator error.Physiologic causes of migration may result from temporary dysmorphism of the ivc including bending, coughing or valsalva maneuvers resulting in dislodgment of the filter.Some studies suggest that strenuous physical activity and increased intra-abdominal pressure can lead to migration of ivc filters.It is unknown if the tilt contributed to the reported perforation.A review of the ifu notes vessel damage such as intimal tears and perforation as procedural complications related to ivc filters.Studies have noted a greater than 80% perforation rate overall, with all filters imaged after 71 days from implantation revealing some level of perforation.Clinical factors that may have influenced the event include the patient¿s pre-existing co-morbidities, pharmacological issues and lesion characteristics.Given the limited information available for review, there is nothing to suggest that a malfunction in the design and manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.Please note that this is the initial report for this product.
|
|
Manufacturer Narrative
|
As reported a patient underwent placement of a trapease vena cava filter.The information provided indicated that the filter subsequently malfunctioned and caused filter fracture, perforation, migration and tilt.The patient reported becoming aware of perforation of filter struts outside the inferior vena cava (ivc), tilting and fracture of the filter, approximately nine years and eleven months post implant.According to the implant record the indication for the filter implant was pulmonary embolus (pe) and non-obstructing right popliteal vein thrombosis.The filter was placed via the right femoral vein and deployed just below the renal veins.The patient tolerated the procedure well.The product was not returned for analysis.A review of the device history record revealed no anomalies during the manufacturing and inspection processes that can be associated with the reported complaint.The trapease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the inferior vena cava (ivc) for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pe where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pe where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Without images or procedural films for review, the reported filter tilt, fracture, migration and perforation events could not be confirmed, nor an exact cause determined.Ivc filter tilt has been associated with practitioner technique and/or vessel anatomy, specifically asymmetry and tortuosity.Additionally, the timing and mechanism of the events are unknown.The instructions for use (ifu) states that filter fracture is a potential complication of vena cava filters.Anatomic locations that create concentrated stress points from filter deformation (for example, deployment at apex of scoliosis, overlapping of either of the renal ostia, or placement adjacent to a vertebral osteophyte) may contribute to fracture of a particular filter strut.Ivc filter migration is a known potential adverse event associated with all ivc filter implants and is listed in the ifu as such.Possible causes for filter migration include mega cava, wire entrapment during central venous catheter placement, ¿sail¿ effect (cranial migration) of large clot burden within the filter, mechanical device failure, and operator error.Physiologic causes of migration may result from temporary dysmorphism of the ivc including bending, coughing or valsalva maneuvers resulting in dislodgment of the filter.Some studies suggest that strenuous physical activity and increased intra-abdominal pressure can lead to migration of ivc filters.It is unknown if the tilt contributed to the reported perforation.A review of the ifu notes vessel damage such as intimal tears and perforation as procedural complications related to ivc filters.Studies have noted a greater than 80% perforation rate overall, with all filters imaged after 71 days from implantation revealing some level of perforation.Clinical factors that may have influenced the event include the patient¿s pre-existing co-morbidities, pharmacological issues and vessel characteristics.Given the limited information available for review, there is nothing to suggest that a malfunction in the design and manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
|
|
Event Description
|
Per the implant records, the patient was reported to have a preoperative diagnosis of pulmonary embolus (pe) and non-obstructing right popliteal vein thrombosis.The patient¿s right groin was prepped and draped in a sterile manner.The right femoral vein was punctured with a needle and a guidewire was introduced under fluoroscopy towards the superior vena cava.A sheath and dilator were introduced for an inferior vena cavagram, which identified the size of the inferior vena cava (ivc) and the level of the renal veins.The filter was delivered just below the renal veins.The patient tolerated the procedure well.According to the information received in the patient profile form (ppf), the patient reports perforation of filter struts outside the inferior vena cava (ivc), tilting and fracture of the filter, becoming aware of these events approximately nine years and eleven months after the filter implantation.
|
|