| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/24/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Report source : foreign : (b)(6).Customer has not indicated whether the product will be returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that there was needle breakage during use.Attempts have been made and additional information on the reported event is unavailable at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.D10 : part cm-9011 quattro suture passer needle lot 79862-1.Visual examination of the returned product identified that the product was returned with significant signs of use with the ribbon tip bent and fractured.The shaft and plastic tab show no signs of damage or deformity.The ribbon tip is in place and the weld holding the ribbon in place is intact.The tip of the ribbon is bent severely and nearly folded in half.The prongs have fractured off, leaving a straight edge.Medical records were not provided.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
Another device was used to complete the procedure, no product fell in patient.No further event information available at the time of this report.
|
| |
|