MAUDE Adverse Event Report: BIOMET 3I FULL OSSEOTITE® TAPERED CERTAIN® IMPLANT 3.25 X 11.5MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I FULL OSSEOTITE® TAPERED CERTAIN® IMPLANT 3.25 X 11.5MM; DENTAL IMPLANT Back to Search Results
Model Number IFNT3211
Medical Device Problem Codes Loss of Osseointegration (2408); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Inflammation (1932); Pain (1994); Osteolysis (2377)
Date of Event 04/28/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).Patient identifier: (b)(6).A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
The doctor indicated loss integration at tooth site #12 due to peri-implantitis.There was no adverse reactions after implantation.The patient felt the pain around the implant.After examination, the clinician found that there was inflammation around the implant and the implant was already loose.The implant was removed by the clinician.
 
Additional Manufacturer Narrative
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: h6: type of investigation code was added: 3331.Dhr review was completed for the subject lot number (2016090577).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Sterilization record (op# (b)(4) was reviewed and verified to have passed all sterilization activities with no issues or nonconformities identified.Complaint history review was performed for the reported lot number (2016090577) for similar events and no other complaint was identified.As documented in the summary investigations, contributing factors for the reported event likely exist outside of zimmer biomet control, including those related to patient biological factors/condition and surgical technique.Based on the summary investigations, no malfunction occurred upon investigation.The reported events remain non-verifiable as they are a medical condition.
 
Event or Problem Description
No additional event information at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FULL OSSEOTITE® TAPERED CERTAIN® IMPLANT 3.25 X 11.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key12681106
Report Number0001038806-2021-02038
Device Sequence Number3983858
Product Code DZE
UDI-Device Identifier00844868004967
UDI-Public(01)00844868004967(17)211007(10)2016090577
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K063341
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/07/2021
Device Model NumberIFNT3211
Device Catalogue NumberIFNT3211
Device Lot Number2016090577
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2021
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/08/2021
Supplement Date Received by Manufacturer03/07/2022
Initial Report FDA Received Date10/22/2021
Supplement Report FDA Received Date03/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/07/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age36 YR
Patient SexMale
-
-