DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problem
Power Problem (3010)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/29/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Testing of actual/suspected device(10): a getinge field service engineer (fse) was dispatched to evaluate this unit.The fse replaced the on/ off switch.The fse completed all pneumatic, safety, functionality, and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that the cs300 intra-aortic balloon pump (iabp) had power switch failure.It is unknown the circumstances under which the event occurred.It is also unknown if there was a patient involvement.However, there was no adverse event reported.
|
|
Manufacturer Narrative
|
Updated fields: b4, g4, g7, g8, h2, h4, h6(investigation type, investigation findings & investigation conclusions), h10, h11.Corrected fields: d5, g1(contact person), h6(clinical code).
|
|
Event Description
|
N/a.
|
|