| Model Number |
27033AA |
| Medical Device Problem Code |
Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The unit in question was requested for return.The investigation is anticipated, but not yet begun.
|
| |
|
Event or Problem Description
|
|
As per a manufacturer incident report we received from the factory in germany: it was reported that during patient treatment the quality of view was too low to work with the optics.The procedure had to be abandoned due to a missing replacement unit.(mfr's.Internal complaint #(b)(4)).
|
| |
|
Additional Manufacturer Narrative
|
|
Upon the evaluation (dated (b)(6) 2021) it could be confirmed that moisture was inside the lens.Moisture was able to penetrate the scope due to mechanical damage to the lens.The most probable root cause is that the lens was damaged due to unanticipated mechanical impact.According to the corresponding instructions for use (mcx620_en_v1.0_ifu_ce-mdr, chapter 3.2), it is written that all components must be checked before each use.The customer will be informed in a separate letter.
|
| |
|