| Catalog Number |
321-25-01 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/04/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Pending evaluation.
|
| |
|
Event or Problem Description
|
|
As reported, while reaming the glenoid with the reverse pilot starter reamer, the pilot tip portion of the reamer broke off in the male patient¿s glenoid.A midas was used to bur out a portion, however, the piece seemed to go deeper into glenoid and surgeon is unsure the entire piece was removed, a portion is believed to be left in patient.Patient was last known to be in stable condition following the event.Device is to be returned.
|
| |
|
Additional Manufacturer Narrative
|
|
(h3) the broken device reported was likely the result of applying a bending moment to the reamer during use, which led to brittle fracture of the pilot tip fracture.(h9) if action reported to fda under 21 usc 360i(f), list correction/removal reporting number: z-2663-2017, z-2664-2017, z-2665-2017, z-2666-2017, z-2668-2017, z-2668-2017.
|
| |
|