| Model Number |
71675208 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
10/05/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).Postal code (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported after procedure, that a intertan 11.5mm x 18cm 130d and their concomitants lag/com screw kit 110/105 and int hex cap scr 5.0mm x 40mm broke beneath the inferior compression screw hole inside the patient.Patient status is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation and the reported event could not be confirmed.The contribution of the device to the reported event could not be corroborated.The clinical/medical investigation concluded that, the provided complaint form supports the complaint.However, smith and nephew has not received the device/adequate clinical documentation to fully evaluate the complaint.Therefore, the root cause of the reported failure could not be confirmed nor concluded.The impact to the patient beyond that which has already been reported cannot be determined.Should any additional relevant medical information be provided, this complaint would be re-assessed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.The instructions for use documents revealed this failure mode was previously identified.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include but not limited to traumatic injury or surgical technique.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Event or Problem Description
|
|
It was reported that, after internal fixation surgery was performed a trigen intertan 11.5mm x 18cm 130 degree nail broke inside the patient just beneath the inferior compression screw hole.It is unknown how this issue was resolved.Current patient's health status is unknown.
|
| |
|