| Model Number |
42866 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/03/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient weight not available from the site.No parts have been received by the manufacturer for evaluation.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a strata valve.Six months after the aneurysm operation, hydrocephalus occurred, and vp surgery was performed in (b)(6) 2021.During the operation, after the valve was connected to the catheter, the reservoir was pressed and the patency check was performed.It was found that the flood-resistant suction chamber of the shunt valve was leaking.The health care professional replaced it with the same model valve that had been stocked by the site.There was no reported impact on the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned valve was patent.The valve met the requirements for valve flux testing, siphon control, reflux, pressure/flow and pre-implantation testing.A tear was observed on the delta chamber of the valve.The catheters associated with the valve were not returned.Due to the presence of the tear, the valve was not able to meet requirements of leakage testing.It is unknown how this damage may have occurred.Ifu cautions, ¿improper use of instruments in the handling or implantation of shunt products may result in the cutting, slitting or crushing of components.Such damage may lead to loss of shunt integrity, and necessitate premature surgical revision of the shunt system.¿ medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received reported that there were no environmental/external/patient factors that may have led or contributed to the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|