Due to character/space constraints within the h10/11 narrative/corrected field, all of the investigation information could not be included and is therefore being added as a separate attachment and should be referenced for complete investigation summary and conclusions.H6: updated investigation conclusion: d15: cause not establish.H6: health effect impact code: f1903: explant.H6: medical device component: g04088: membrane.The investigation has been completed.In the absence of additional information or medical records from the complainant, this event file will be closed with the information provided.Conclusion: implant #8 gore® synecor intraperitoneal biomaterial it should be noted that the [gore® synecor intraperitoneal biomaterial instructions for use include warnings and addresses the following adverse reactions among others: ¿possible adverse reactions with the use of any tissue deficiency prosthesis may include, but are not limited to, adhesions and related harms, bowel obstruction, contamination, defect recurrence and related harms, dysphagia, erosion or extrusions and related harms, exposure or protrusions and related harms, fever, fistula, gerd recurrence, infection, irritation or inflammation, mesh migration, mesh shrinkage, pain, parathesis, seroma or hematoma and related harms, tissue ischemia, wound dehiscence and additional intervention including surgery.The instructions for use further warn: ¿strict aseptic techniques should be followed.When operative infection is suspected, dissection of involved tissues should be considered.If an infection develops, it should be treated aggressively.An unresolved infection may require removal of the material.¿ the instructions for use also warn, ¿when using this device as a permanent implant and exposure occurs, treat to avoid contamination, or device removal may be necessary.¿ procedure and specific patient factors may contribute to or cause infection, leading to contamination, exposure, lack of incorporation and/or seeding of device.Procedure related factors may include adherence to clinical guidelines on infection risk management, contamination of device prior to or during implant, and post-operative persistent/symptomatic seroma and wound management.Patient risk factors may include diabetes, smoking, age, malnutrition, immunosuppressive therapy, post-operative instruction noncompliance, and hygiene.The instructions for use state: ¿the use of absorbable sutures may lead to inadequate anchoring of gore® synecor intraperitoneal biomaterial to the host tissue and necessitate reoperation.For best results, use monofilament sutures.¿ there is insufficient information available for gore to reasonably draw conclusions related to aspects of the event, therefore conclusion code "d15: cause not established" is being used.Insufficient information may include limited or missing relevant medical records, involvement of multiple implanted devices (including non-gore devices) in the field of treatment, patient non-compliance, and/or a general lack of available detail or specificity related to an adverse event and/or device.As with any surgical procedure, there are always risks of complications for surgical repair of hernias and soft tissue deficiencies, with or without mesh.These may include but are not limited to, adhesions and related harms, bleeding, bowel obstruction, compromised device biocompatibility, contamination which may lead to patient harms, device damage, dysphagia, erosion or extrusion and related harms, exposure or protrusion and related harms, fever, fistula, gerd recurrence, defect recurrence and related harms, ileus, increased procedure time and related harms, irritation or inflammation, infection, mesh migration, mesh shrinkage, pain, paresthesia, perforation, revision /re-intervention, seroma or hematoma and related harms, tissue ischemia, wound complications and wound dehiscence and additional intervention including surgery.Many of the potential complications are associated with the patient¿s underlying disease progression, co-morbidities, additional medical history and/or other surgical procedures.The above inherent risks are typically detailed in standard informed consent documents.Based upon the information received, status of the device is unable to be confirmed and therefore not available for evaluation.Review of the manufacturing and sterilization records verified that the lot met all pre-release specifications.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|