• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. AUTOCAT2 WAVE; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. AUTOCAT2 WAVE; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Catalog Number IAP-0500
Device Problems Mechanical Problem (1384); Pressure Problem (3012)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/24/2021
Event Type  malfunction  
Manufacturer Narrative
Qn#: (b)(4).
 
Event Description
There was no patient involvement.It was reported that a failure was found during preventive maintenance of the device.The pump displayed the error "high baseline" while the pump was displaying an irregular behavior and sound in recalling and providing the gas to the balloon simulator.It was noted that the pcs mounted on the pump was in perfect condition, so it has not been replaced.As a result, the pump assembly (without the pcs) has been replaced and the machine was put back in service.
 
Manufacturer Narrative
Qn# (b)(4).Teleflex received the device for investigation.The reported complaint of "high baseline alarm" is confirmed.The pump alarmed for "high baseline" during the complaint investigation.An excessive noise was noted from the pump assembly consistent with a motor/lead screw malfunction.The root cause of this complaint is undetermined.A device history record (dhr) review was conducted for the lot number/serial number with no relevant findings.The device passed all manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revised risk.This will be monitored for any developing trends.
 
Event Description
There was no patient involvement.It was reported that a failure was found during preventive maintenance of the device.The pump displayed the error "high baseline" while the pump was displaying an irregular behavior and sound in recalling and providing the gas to the balloon simulator.It was noted that the pcs mounted on the pump was in perfect condition, so it has not been replaced.As a result, the pump assembly (without the pcs) has been replaced and the machine was put back in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOCAT2 WAVE
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
kaylia pen
16 elizabeth drive
chelmsford, MA 01824
MDR Report Key12684286
MDR Text Key278105111
Report Number3010532612-2021-00351
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K060309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 09/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberIAP-0500
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/29/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/19/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-