Catalog Number IAP-0500 |
Device Problems
Mechanical Problem (1384); Pressure Problem (3012)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/24/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Qn#: (b)(4).
|
|
Event Description
|
There was no patient involvement.It was reported that a failure was found during preventive maintenance of the device.The pump displayed the error "high baseline" while the pump was displaying an irregular behavior and sound in recalling and providing the gas to the balloon simulator.It was noted that the pcs mounted on the pump was in perfect condition, so it has not been replaced.As a result, the pump assembly (without the pcs) has been replaced and the machine was put back in service.
|
|
Manufacturer Narrative
|
Qn# (b)(4).Teleflex received the device for investigation.The reported complaint of "high baseline alarm" is confirmed.The pump alarmed for "high baseline" during the complaint investigation.An excessive noise was noted from the pump assembly consistent with a motor/lead screw malfunction.The root cause of this complaint is undetermined.A device history record (dhr) review was conducted for the lot number/serial number with no relevant findings.The device passed all manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revised risk.This will be monitored for any developing trends.
|
|
Event Description
|
There was no patient involvement.It was reported that a failure was found during preventive maintenance of the device.The pump displayed the error "high baseline" while the pump was displaying an irregular behavior and sound in recalling and providing the gas to the balloon simulator.It was noted that the pcs mounted on the pump was in perfect condition, so it has not been replaced.As a result, the pump assembly (without the pcs) has been replaced and the machine was put back in service.
|
|
Search Alerts/Recalls
|