• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH INC LANCET-100; LANCET, BLOOD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH INC LANCET-100; LANCET, BLOOD Back to Search Results
Model Number LCT, TP GNP 30G 100CT50/CASE DK BL
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/24/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Lancets were returned for evaluation.Investigation in process.Note: manufacturer contacted customer in a follow-up call on 05-oct-2021 to ensure the replacement products resolved the initial concern-able to establish contact with customer.Customer doesn¿t have the new product with him but he stated that he is not having issues with the new box of lancets but he also stated that even tough we recommended him to cease the use of the lancets that he alleged were defective he was using more lancets from that lot # and they were working fine and he will send us back the lancets that he found with the issue (needle came out with the protective cap) and the lancets that he has left in that box.
 
Event or Problem Description
Consumer reported complaint for the 30g lancets.Customer stated the needle came off when he twisted and removed the protective cap.Customer stated he had tried two more lancets from the same box and the needle had come off as well.The package had not been open or damaged when received by the customer.This was the first time the customer had used the product out of this package.The customer feels well and did not report any symptoms.The customer did not claim they were injured while using the lancets and no medical attention associated with the use of the product was reported.
 
Additional Manufacturer Narrative
Sections with additional information as of 23-nov-2021: h3: was the device evaluated by the manufacturer.H6: updated fda¿s type, findings and conclusions codes.H10: lancets were returned for evaluation.Lancets were visually inspected: needle did not come out upon removal of lancet cap, no defect found.Returned product was forwarded to supplier quality to contact manufacturer.Manufacturer investigation completed and no abnormalities observed on returned product, retain samples, and review batch record.Most likely underlying root cause: mlc-061: improper use/mishandle by end user.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LANCET-100
Common Device Name
LANCET, BLOOD
Manufacturer (Section D)
TRIVIDIA HEALTH INC
2400 nw 55th court
fort lauderdale FL 33309
Manufacturer Contact
karen devincent
2400 nw 55th court
fort lauderdale, FL 33309
MDR Report Key12685741
Report Number1000113657-2021-00666
Device Sequence Number3707970
Product Code FMK
UDI-Device Identifier00087701426056
UDI-Public087701426056
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 11/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/20/2025
Device Model NumberLCT, TP GNP 30G 100CT50/CASE DK BL
Device Lot Number201039NM
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2021
Is the Reporter a Health Professional? No
Distributor Facility Aware Date09/24/2021
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/24/2021
Supplement Date Received by Manufacturer11/02/2021
Initial Report FDA Received Date10/22/2021
Supplement Report FDA Received Date11/23/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-