| Model Number |
L221 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/29/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Following confirmation of explant, this event will be further updated.
|
| |
|
Event or Problem Description
|
|
It was reported that during a routine follow-up, this device exhibited safety mode operation.No resulting adverse patient effects were reported and all attempts to obtain resolution were unsuccessful.Our information to date is suggestive that this device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.A high current drain was detected, but the battery still supported full device function.Detailed analysis confirmed the identified high current drain was a result of compromised capacitors.This resulted in the observed premature battery depletion.It was determined that the capacitors were compromised due to the presence of excess hydrogen in the device case.Boston scientific issued a field safety notice regarding a subset of devices in the accolade pacemaker family that have a potential of exhibiting this behavior.This particular device is included in the hydrogen induced premature depletion advisory population.
|
| |
|
Event or Problem Description
|
|
It was reported that during a routine follow-up, this device exhibited safety mode operation.No resulting adverse patient effects were reported and all attempts to obtain resolution were unsuccessful.The device was later explanted and returned for analysis.
|
| |
|