Model Number VERSION 1.0
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Perforation (2001)
Date of Event 09/27/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.The medical device manufacturer and manufacturing location for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.(expiry date: 07/2023) device pending return.
 
Event or Problem Description
It was reported that during a procedure, the device allegedly caused perforation in the wall of the distal popliteal artery.The patient was reportedly stable.
 
Event or Problem Description
It was reported that during a recanalization procedure, the device allegedly caused perforation in the wall of the distal popliteal artery.The patient was reportedly stable.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was returned to the manufacturer for evaluation.A catheter was physically investigated.The catheter was blocked with thrombotic material from patients body.After the flushing normal aspiration was restored.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: d4 (expiry date: 07/2023), g3.H11: h6 (method, result, conclusion).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAREX
Common Device Name
THROMBECTOMY & ATHERECTOMY
Manufacturer (Section D)
UNKNOWN
BR 
Manufacturer (Section G)
UNKNOWN
BR  
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key12687996
Report Number3008439199-2021-00172
Device Sequence Number4051457
Product Code MCW
UDI-Device Identifier07640142810575
UDI-Public(01)07640142810575
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K172315
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVERSION 1.0
Device Catalogue Number80236
Device Lot Number200992
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/27/2021
Supplement Date Received by Manufacturer03/18/2022
Initial Report FDA Received Date10/25/2021
Supplement Report FDA Received Date03/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexMale
-
-