• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SANMINA -SCI SYSTEMS O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SANMINA -SCI SYSTEMS O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number BI70002000
Medical Device Problem Code Device Sensing Problem (2917)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A medtronic representative went to the site to test the equipment.Testing revealed that no failure was found.The system then passed the system checkout and was found to be fully functional.B01, c19, d14 applicable codes.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding a navigation device being used for a unknown procedure.It was reported that the image acquisition system (ias) showed " door open " but it closes.It cannot open the door by the pendant and emergency door override.No impact on patient outcome.The inner gantry door covers was uncovered and found the dinguses in the wrong position of the gear tooth.It was adjusted to the correct position.
 
Event or Problem Description
Conflicting information was received.It was reported, that no patient was present, and that there was no impact to the patient.It was unclear whether, or not a patient was present".
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Common Device Name
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SANMINA -SCI SYSTEMS
km 15.5 no. 29, plant 06
carr. chapala-guadalajara,jal 45640
MX  45640
Manufacturer (Section G)
SANMINA -SCI SYSTEMS
km 15.5 no. 29, plant 06
carr. chapala-guadalajara,jal 45640
MX   45640
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key12691611
Report Number3006544299-2021-00396
Device Sequence Number4103565
Product Code OWB
UDI-Device Identifier00763000355609
UDI-Public00763000355609
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K200074
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBI70002000
Device Catalogue NumberBI70002000
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2021
Supplement Date Received by Manufacturer11/02/2021
Initial Report FDA Received Date10/25/2021
Supplement Report FDA Received Date12/01/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/20/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-