• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE, INCORPORATED NUVASIVE RELINE SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE, INCORPORATED NUVASIVE RELINE SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number 11102255
Medical Device Problem Code Separation Problem (4043)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/16/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product has been received along with radiographs provided confirmed the complaint.The patients post op physical activity levels are unknown.It is also unknown if the patient experienced a fall or other accident.Examination of the returned connector found the closed lock screw to be cross threaded and not securing the rod and is considered the root cause of the post-operative rod separation.No additional investigation required.Label review "potential adverse events and complications - as with any major surgical procedures, there are risks involved in orthopedic surgery.Infrequent operative and postoperative complications that may result in the need for additional surgeries.Potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s) loss of fixation." "warnings, cautions and precautions: all lock screws should be final-tightened with the counter-torque and torque t-handle.Do not final-tighten through compression instruments (e.G.C/d rack and figure 8 compressor) in the set, as the rod may not be able to normalize to the tulip.Be cautious not to over compress or distract as you can loosen the screws in the spine and potentially pull out the screw.The bulleted portion of the nose of the rod and the faceted portion of the rod (where the inserter locks down on the rod) must extend fully outside of the most inferior or most superior tulip on the construct.The set screw cannot be locked down on this unusable portion of the rod, as this may compromise the stability of the construct." "patient education: preoperative instructions to the patient are essential.The patient should be made aware of the limitations of the implant and potential risks of the surgery.The patient should be instructed to limit postoperative activity, as this will reduce the risk of bent, broken or loose implant components.The patient must be made aware that implant components may bend, break or loosen even though restrictions in activity are followed." "compatibility: all implants should be used only with the appropriately designated instrument reference surgical technique." "pre-operative warnings: 3.Care should be used in the handling and storage of the reline implants.The implants should not be scratched or damaged.Implants and instruments should be protected during storage and from corrosive environments.5.Care should be used during surgical procedures to prevent damage to the device(s) and injury to the patient.".
 
Event or Problem Description
On (b)(6) 2021 a patient underwent a posterior lumbar interbody fusion procedure from t12 to l2.During a post-op follow up on and unknown date it was discovered via radiograph that the rod slipped and displaced from the connector.On (b)(6) 2021 a revision occurred were both implanted connectors and the rods were replaced.The patient is recovering well post revision.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE RELINE SYSTEM
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
NUVASIVE, INCORPORATED
7475 lusk boulevard
san diego CA 92121
MDR Report Key12691667
Report Number2031966-2021-00144
Device Sequence Number4165760
Product Code NKB
UDI-Device Identifier00887517596185
UDI-Public887517596185
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K143684
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number11102255
Device Lot NumberTU7914
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2021
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/30/2021
Initial Report FDA Received Date10/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age73 YR
-
-