| Catalog Number |
5C4456 |
| Medical Device Problem Code |
Unsealed Device Packaging (1444)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/30/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that pouches of an unspecified quantity of flexicaps were opened.This was observed before use of the device for peritoneal dialysis therapy.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
One (1) carton and twenty-eight (28) unopened pouched units were received for evaluation.A visual inspection was performed, and it was noted one (1) of the pouched units was open at the side of the pouch.The reported condition was verified.The cause of the reported condition could not be determined; however, the probable cause was a shipping related issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|