| Catalog Number |
46538BER |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/23/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that during a tunneled dialysis catheter placement procedure the catheter tip detached within the patient.The clinician had acquired antegrade access within the patient's internal jugular vein and while attempting to cannulate the patient superior vena cava [svc] with the guidewire the catheter tip detached.The clinician externalized the foreign body together with the guidewire liberating the svc.No additional interventional procedure was necessary.Another gw and catheter was prepped to successfully complete this procedure.Only a small delay in procedure to report.No additional patient consequences to report.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was returned for evaluation.The complaint was confirmed and the root cause is attributed to clinical use and excessive force during use.A review of the device history and complaint database could not be performed since the lot number was not provided.
|
| |
|