| Model Number |
3311 |
| Medical Device Problem Codes |
Failure to Power Up (1476); Power Problem (3010)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/16/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that upon testing by pulmolink, both cables were found to have suspected fractures within the plug/strain relief at the sensor end of the cable which caused intermittent signal transmission when gently flexed.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: additional information added to h6 and h10.This remediation mdr was generated under protocol (b)(4) as a result of warning letter cms# (b)(4) a product sample was received for evaluation.Visual and functional testing were performed.Both cables were found to have suspected fractures within the plug/strain relief at the sensor end of the cable which caused intermittent signal transmission when gently flexed.The root cause of the reported issue was found to be cable interruption at the socket.Complaint investigation was brought to the attention of the manufacturing engineers back in july 2020.
|
| |
|