DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problems
Inability to Auto-Fill (1044); Leak/Splash (1354); Suction Problem (2170)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/05/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Testing of actual/suspected device(10): a getinge field service engineer (fse) evaluated the iabp unit and confirmed the issue.The fse found that k7 would cycled on/off when the iabp unit sat idle and it would fail the k6 - k8 leak test in diagnostics.To fix the issue, the fse replaced the drive manifold assembly and performed all calibration, functional and safety checks to meet factory specifications.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) giving low suction error and also leak failure.No patient harm, serious injury or adverse event was reported.
|
|
Event Description
|
It was reported by the customer that the cs300 intra-aortic balloon pump (iabp) was giving a low suction error, and also a leak failure.It is unknown the circumstances under which the event occurred.No patient harm, serious injury or adverse event was reported.
|
|