LUMENIS LTD. ACUPULSE 40 WATT; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number ACUPULSE 40 WATT
Medical Device Problem Code Failure to Deliver Energy (1211)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/02/2021
Type of Reportable Event Malfunction
Event or Problem Description
A user facility reported that a lumenis acupulse 40w co2 laser system presented error 66 - scanner communication error, on its display.Unable to proceed with the laser, the procedure was cancelled.No report of injury was received, and no report was received alleging the device malfunction caused or contributed to any change in the patient's condition.
 
Additional Manufacturer Narrative
Lumenis investigated the reported event by contacting the user facility directly for additional information.Despite reasonable attempts, eight (8) by phone and by email, no additional information was provided.The question if the procedure was canceled prior to anesthetizing the patient remained unanswered.A review of the subject device dhr confirmed that the subject device was manufactured on 10-aug-2014 and installed at the customers site on (b)(6) 2014.A lumenis service engineer visited the site twenty three (23) days after the reported event and examined the laser system.The engineer confirmed the issue as reported and replaced the scanner controller board, performed all system calibrations and tests.The device was found to have met lumenis specifications and ready for use.A review of system risk files (rd-1148040 rev t) revealed risk #10.1 " device malfunction" which have the potential to lead to adverse effects of alternative treatments".The risk likelihood has been quantified and found to be remote, and the risk has been characterized and documented as acceptable within full risk assessment.In this case, the procedure was cancelled, though there is no indication that the patient was under anesthesia, in an abundance of caution, lumenis is reporting this malfunction.According to the gso expert based on the age of the system, wear could have likely played a role in the failure.Therefore, no additional corrective or preventive action is determined necessary.Lumenis is closing this complaint but will continue to monitor this failure mode; complaint trending will continue to monitor per global complaint handling sop (doc no.(b)(4)) and per post marketing surveillance procedure (doc no.(b)(4)).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACUPULSE 40 WATT
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
LUMENIS LTD.
6 hakidma st.
po box 240
yokneam, 20692
IS  20692
MDR Report Key12695208
Report Number3004135191-2021-00066
Device Sequence Number4445052
Product Code GEX
UDI-Device Identifier07290109140315
UDI-Public07290109140315
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K100415
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberACUPULSE 40 WATT
Device Catalogue NumberGA-1000000
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/02/2021
Initial Report FDA Received Date10/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-