• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. 4.5X32MM PERIPHERAL SCREW, NON-LOCKING; SHOULDER PROSTHESIS, REVERSE CONFIGURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. 4.5X32MM PERIPHERAL SCREW, NON-LOCKING; SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Model Number 4.5X32MM PERIPHERAL SCREW, NON-LOCKING
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/12/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
On 09/10/2021, it was reported through the surgical outcome system that a revision surgery is required due to the loosening of the prosthetic joint.No additional information provided.Additional information requested.Additional information provided 10/5/2021: the date of primary procedure was on (b)(6) 2019 for a reverse total shoulder replacement.This procedure took place at beachwood medical center, arthrex account (b)(4).The date of the revision surgery was on (b)(6) 2019 for a poly swap and on another on (b)(6) 2020 for a poly swap and glenosphere exchange.These revision(s) took place at beachwood medical center, arthrex account (b)(4).The part and lot numbers of all arthrex parts that were explanted during the revision(s) are as followed: on 12/12/2019: insert (ar-9503-3633-6, lot 18.02045 ((b)(6) 2024)), peripheral screw (ar-9562-32nl, lot 2019001013, ((b)(6) 2024)) on 2/4/2020: glenosphere (ar-9564-2433-lat, lot 19.00014 (2024-04-30)), insert (ar-9503-3633-3, lot 19.00010 ((b)(6) 2024)), unsure which specific screw replaced.The part and lot numbers of all arthrex parts implanted during the revision(s) are as followed: on (b)(6) 2019: univers revers mgs peripheral screw (qty 2), univers revers mgs combo humeral insert, univ.Rev.Spacer on (b)(6) 2020: univ.Rev mgs glenosphere, univ.Rev.Humeral insert, univ.Rev.Mgs peripheral screw, (qty.1).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5X32MM PERIPHERAL SCREW, NON-LOCKING
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key12696668
Report Number1220246-2021-03832
Device Sequence Number20375095
Product Code PHX
UDI-Device Identifier00888867296718
UDI-Public00888867296718
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K173900
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2019
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 10/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Model Number4.5X32MM PERIPHERAL SCREW, NON-LOCKING
Device Catalogue NumberAR-9562-32NL
Device Lot Number2019001013
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/05/2021
Initial Report FDA Received Date10/26/2021
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-