| Model Number |
DTMB1D4 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Electromagnetic Interference (1194)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/21/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: 511212 lead implanted: (b)(6) 2017.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a single high rate non-sustained episode that showed non-physiological noise on the atrial and ventricular electrograms (egm).It was noted this was likely due to electromagnetic interference.The patient believed they were using a lawn mower at the time.The cardiac resynchronization therapy defibrillator (crt-d) remains in use.No patient complications have been reported as a result of this event.
|
| |
|
Event or Problem Description
|
|
Additional information received indicated electromagnetic interference was confirmed.The patient was using an old lawnmower at the same time as the short v-v episode.There was no issue with the leads and function has remained normal.No reprogramming or intervention was done.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|