C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY® URETERAL STENT WITH NICORE¿ GUIDEWIRE Back to Search Results
Model Number 776624
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that the patient experienced edema, stone formation, ureteral reflux, stent dislodgement, migration, stent encrustation, infection, pain and discomfort while using the bard ureteral stent.No medical intervention was reported.
 
Event or Problem Description
It was reported that the patient experienced edema, stone formation, ureteral reflux, stent dislodgement, migration, stent encrustation, infection, pain and discomfort while using the bard ureteral stent.No medical intervention was reported.
 
Additional Manufacturer Narrative
The reported event is inconclusive since no sample was returned.A potential root cause for this event could be, "material selection" or "part geometry".The device was used for treatment purposes, however, it is unknown if the device had met relevant specifications or contributed to the reported event.The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "potential complications associated with retrograde/antegrade positioning of indwelling ureteral stents include the following: edema; stone formation; peritonitis; extravasation; ureteral reflux; stent dislogdgement, fragmentation, migration, occlusion; fistula formation; loss of renal function; hemorrhage; pain/discomfort; stent encrustation; hydronephrosis; perforation of kidney, renal pelvis, ureter and/or bladder; ureteral erosion; infection; urinary symptoms." "with any ureteral stent, migration is a possible complication which could require medical intervention for removal.Selection of too short a stent may result in migration." "determine the proper stent length for the patient.This is generally calculated from the baseline pyelogram." h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® INLAY® URETERAL STENT WITH NICORE¿ GUIDEWIRE
Common Device Name
URETERAL STENT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key12698355
Report Number1018233-2021-06665
Device Sequence Number11896730
Product Code FAD
UDI-Device Identifier10801741014267
UDI-Public(01)10801741014267
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K983498
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number776624
Device Catalogue Number776624
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/06/2021
Supplement Date Received by Manufacturer03/28/2022
Initial Report FDA Received Date10/26/2021
Supplement Report FDA Received Date04/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-