• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH LOCKING SCREW VARIAX 2 T10, FULL THREAD, 3.5MM / L26MM; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH LOCKING SCREW VARIAX 2 T10, FULL THREAD, 3.5MM / L26MM; PLATE, FIXATION, BONE Back to Search Results
Model Number 657326S
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device will not be returned.If additional information becomes available, it will be provided in a supplemental report.Device was discarded by hospital.
 
Event or Problem Description
It was reported a locking failure occurred.It is possible that the allowable angle of locking has been exceeded.
 
Event or Problem Description
It was reported a locking failure occurred.It is possible that the allowable angle of locking has been exceeded.
 
Additional Manufacturer Narrative
The reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.A potential non-conformity report was nonetheless initiated to address this event.The comprehensive root cause analysis of the potential ncr included a surface and microstructure analysis as well as nano-indent hardness measurement of variax1 and variax2 screw samples by the (b)(6) on behalf of stryker.Further internal investigation efforts were focusing on the anodization color difference between variax1 and variax2 screws.Significant differences between the screws which could result in the reported complaints have not been revealed in either fraunhofer or stryker investigations.Excessive in-house handling and bench testing with variax2 screws from various manufacturing batches showed that locking is achieved as intended.The desired increase in torque indicating to the customer that the screw is locking in the plate, can be confirmed.Considering the information given and based on the above investigations a root cause of the reported event could not be determined.A review of the device history for the reported lot did not indicate any abnormalities.A review of the labeling did not indicate any abnormalities.If any further substantial information is provided, the investigation report will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW VARIAX 2 T10, FULL THREAD, 3.5MM / L26MM
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
marilyne chaumont
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key12699085
Report Number0008031020-2021-00440
Device Sequence Number3984388
Product Code HRS
UDI-Device Identifier07613327087024
UDI-Public07613327087024
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K132502
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2022
Device Model Number657326S
Device Catalogue Number657326S
Device Lot NumberL35232
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/01/2021
Supplement Date Received by Manufacturer11/18/2021
Initial Report FDA Received Date10/26/2021
Supplement Report FDA Received Date12/16/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/14/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-