MAUDE Adverse Event Report: STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP SMARTLIFE LARGE ASPETIC HOUSING; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP SMARTLIFE LARGE ASPETIC HOUSING; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Model Number 7126-120-000
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/14/2021
Event Type  malfunction  
Event Description
The user facility reported that the housing broke at the weld during a procedure.The broken housing could cause the non-sterile battery to be exposed to the sterile field.There were no adverse consequences, no medical intervention and no clinically significant delay.
 
Event Description
The user facility reported that the housing broke at the weld during a procedure.The broken housing could cause the non-sterile battery to be exposed to the sterile field.There were no adverse consequences, no medical intervention and no clinically significant delay.
 
Manufacturer Narrative
The reported event was confirmed.A picture was attached at intake.It was visually observed the weld was compromised.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTLIFE LARGE ASPETIC HOUSING
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP
1941 stryker way
portage MI 49002
Manufacturer (Section G)
STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP
1941 stryker way
portage MI 49002
Manufacturer Contact
zach baker
1941 stryker way
portage, MI 49002
2693237700
MDR Report Key12699087
MDR Text Key278407574
Report Number3015967359-2021-02400
Device Sequence Number1
Product Code HWE
UDI-Device Identifier07613327128819
UDI-Public07613327128819
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/26/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7126-120-000
Device Catalogue Number7126120000
Device Lot Number19002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received10/15/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/02/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age68 YR
Patient SexMale
-
-