STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP SMARTLIFE LARGE ASPETIC HOUSING; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Model Number 7126-120-000
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/22/2021
Event Type  malfunction  
Manufacturer Narrative
The reported event was confirmed.A picture was attached at intake.It was visually observed the weld was compromised.
 
Event Description
The user facility reported that the housing broke at the weld during a procedure.The broken housing could cause the non-sterile battery to be exposed to the sterile field.There were no adverse consequences, no medical intervention and no clinically significant delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTLIFE LARGE ASPETIC HOUSING
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP
1941 stryker way
portage MI 49002
Manufacturer (Section G)
STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP
1941 stryker way
portage MI 49002
Manufacturer Contact
zach baker
1941 stryker way
portage, MI 49002
2693237700
MDR Report Key12699090
MDR Text Key278411315
Report Number3015967359-2021-02399
Device Sequence Number1
Product Code HWE
UDI-Device Identifier07613327128819
UDI-Public07613327128819
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/26/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7126-120-000
Device Catalogue Number7126120000
Device Lot Number20153
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/22/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-