• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. VCL CT BRD VIO 18IN 6-0 D/A TG100-8; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VCL CT BRD VIO 18IN 6-0 D/A TG100-8; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Model Number J544G
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/24/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent a bilateral rectus recession procedure on (b)(6) 2021 and suture was used.During the procedure, the suture separated from the needle.Another like device was used to complete the procedure.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested and the following was obtained: how many sutures separated from the needle during use on this this single procedure being reported? two was the procedure completed successfully? how? yes.Continued to open additional sutures.When did the event occur? pre-op or intra-op? intraoperative procedure name and date? bilateral rectus recession (b)(6) 2021 a manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Events reported in 2210968-2021-10348 and 2210968-2021-10349.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 12/2/2021.Additional information: d4, d9, h3, h4, h6.H6 component code: g07002 no device problem found.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational narrative: a sealed box with twelve packets and an opened box with five packets of product code j544g was returned to ethicon inc for analysis with the packaging closed.Upon initial inspection, of the samples, no external damages were observed on the packet.In order to evaluate the conditions of the returned samples, thirteen packets were opened, and no defects were detected.The swage and attachment area was noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related no defects were observed during evaluation.A functional test was performed, and the pull force meets the requirements.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL CT BRD VIO 18IN 6-0 D/A TG100-8
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
avenida de las torres 7125
col salvacar
cuidad juarez
MX  
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
MDR Report Key12699921
Report Number2210968-2021-10349
Device Sequence Number4641386
Product Code GAM
UDI-Device Identifier10705031041073
UDI-Public10705031041073
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K022269
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/05/2021
Device Model NumberJ544G
Device Catalogue NumberJ544G
Device Lot NumberRDMAMT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2021
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/29/2021
Supplement Date Received by Manufacturer11/11/2021
Initial Report FDA Received Date10/26/2021
Supplement Report FDA Received Date12/02/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-