| Model Number |
2545-01-046 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/13/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the patella drill trials spikes are missing.Received tray with patella trials missing the spikes.No surgical delay.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # : (b)(4).Investigation summary : examination of the returned device cannot confirm the breakage condition.However, the evaluation brought forth discrepancies between the drawing and the returned product depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|