PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Model Number 400SMTSFT0506
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report numbers.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the basilar artery using penumbra smart coils (smart coil), penumbra smart coil detachment handles (handle), a non-penumbra microcatheter, a 5f diagnostic catheter, and a guidewire.During the procedure, the physician advanced a smart coil to the target vessel and attempted to detach it using a handle; however, the smart coil failed to detach.Therefore, the smart coil was removed.Subsequently, the same issue occurred with eight other smart coils.It was reported that attempts were made to detach the smart coils with three different handles.The procedure was completed using new smart coils.There was no report of an adverse effect to the patient.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the basilar artery using penumbra smart coils (smart coil), a penumbra smart coil detachment handle (handle), a non-penumbra microcatheter, a 5f diagnostic catheter, and a guidewire.During the procedure, the physician advanced a smart coil to the target vessel and attempted to detach it using the handle; however, the smart coil failed to detach.Therefore, the smart coil was removed.Subsequently, the same issue occurred with five other smart coils.It was reported that attempts were made to detach the smart coils with three different handles.The procedure was completed using new smart coils and a new handle.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Evaluation of the returned smart coil confirmed that the embolization coil was intact with the pusher assembly.Further evaluation revealed that the pet lock was intact on the proximal end of the pusher assembly, and pusher assembly had kinks along its length.The pet lock being intact indicate that there was no proper attempt to detach the coil during the procedure.If the proximal end of the pusher assembly is not properly inserted into a handle prior to a detachment attempt, the pet lock will not separate, and the embolization coil will not detach.The smart coil was attempted to be detached using a demonstration handle but was unsuccessful as the pet lock separated, but the pull wire would not retract out of the pusher assembly ddt, and the embolization coil did not detach.Please note that the following section is being updated based on additional information provided by a penumbra sales representative on 27-jan-2022: 1.Section b.Box 5.Describe event or problem.Please note that the following sections were incorrectly reported on the initial mfr report and are being corrected on this follow-up #01 mfr report:3005168196-2021-02389.1.Section d.Box 4.Lot number, catalog number, expiration date, unique identifier.2.Section h.Box 4.Device manufacture date.This report is associated with mfr report numbers: 1.3005168196-2021-02388, 2.3005168196-2021-02390, 3.3005168196-2021-02391, 4.3005168196-2021-02392, 5.3005168196-2021-02393, 6.3005168196-2021-02394, 7.3005168196-2021-02396, 8.3005168196-2021-02398, 9.3005168196-2021-02399, 10.3005168196-2021-02401, 11.3005168196-2021-02402 h3 other text : placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key12701111
Report Number3005168196-2021-02389
Device Sequence Number6628530
Product Code HCG
UDI-Device Identifier00814548015149
UDI-Public00814548015149
Combination Product (Y/N)Y
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/16/2022
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/02/2026
Device Model Number400SMTSFT0506
Device Catalogue Number400SMTHXSFT0306
Device Lot NumberF106806
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2021
Supplement Date Received by Manufacturer02/14/2022
Initial Report FDA Received Date10/26/2021
Supplement Report FDA Received Date02/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Date Report Sent to FDA01/01/2005
-
-