Model Number A35HPV12080080
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/15/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Physician was attempting to use the fortrex pta balloon catheter during renal access procedure.There was no leak noted and no luer adapter issue.No resistance was encountered when advancing the device.It was reported that the first balloon was filled to the rated burst pressure of 12 in normal use.When the stenosis was found not to be expanded and continued to expand it was filled to the burst pressure of 6 and the balloon ruptured longitudinally.Replaced with a second balloon of the same product/model and lot number immediately.When the rated burst pressure was 7 it ruptured longitudinally again.The balloon did not fragment when it burst.There was no intervention required to remove the burst balloon from the patient.The burst balloon was retrieved from patient's vessel.The device was intact on removal.There was no vessel damage.The procedure was completed by replacing with another brand of balloon immediately.No patient injury reported.
 
Additional Manufacturer Narrative
Product analysis: the fortrex pta balloon was returned to medtronic investigation lab for evaluation.The device returned coiled in a biohazard bag.Device was decontaminated with cidex opa solution soak and tergazyme soak.The fortrex balloon catheter was received with a longitudinal tear in the balloon chamber.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTREX 0.035 OTW PTA BALLOON CATHETER
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key12702968
Report Number2183870-2021-00391
Device Sequence Number1389140
Product Code LIT
UDI-Device Identifier00763000106430
UDI-Public00763000106430
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K142654
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/28/2023
Device Model NumberA35HPV12080080
Device Catalogue NumberA35HPV12080080
Device Lot NumberB016013
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/15/2021
Supplement Date Received by Manufacturer03/03/2022
Initial Report FDA Received Date10/27/2021
Supplement Report FDA Received Date03/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/28/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age43 YR
Patient SexFemale
-
-