| Model Number |
G5A-80A |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device data logs.However, the device was put through extensive testing including the periodic self test, multiple shock testing, and functional testing without duplicating the report.The report was cleared and did not reoccur.An internal inspection of the device found no discrepancies.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device displayed an "ecg fault -501" error message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Search Alerts/Recalls
|