SYNTHES GMBH SAGITTAL SAW ATTACHMENT, LONG, FOR TRS; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 05.001.224 |
Device Problems
Mechanical Jam (2983); Complete Loss of Power (4015)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to component wear.Udi: (b)(4).
|
|
Event Description
|
It was reported from (b)(6) that during service and evaluation, it was determined that the saw attachment device was frozen/would not move.It was noted in the service order that the device did not work.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2021.All available information has been disclosed.If additional information should become available, an additional report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|